INSTRUCTION
MANUAL
HIGH VOLTAGE UNIT
TYPE : HV501
NUCLEONIX SYSTEMS PRIVATE LIMITED
Plot No : 162 A & B, PHASE II, I.D.A.Cherlapally, Hyderabad - 500 051
Ph: 91-040-27263701/30918055, FAX : 27262146
e-mail : [email protected] www.nucleonix.com
File Name
Date
: HV501_MAN
: 20-08-2012
CONTENTS
Sl.No.
Description
Unpacking
CHAPTER I
Introduction
CHAPTER II
Specifications
CHAPTER III
Installation
CHAPTER IV
Operating Instructions
CHAPTER V
Block diagram & Description
CHAPTER VI
Availing of Mainstenance / Calibration Services
and Warranty Clause
IMPORTANT NOTES
1.
POLARITY REVERSAL
1.
If the unit is in operation with a particular POLARITY, reduce the EHT to minimum and
switch off the "EHT ON" switch.
2.
Open the right 'side cover'. (Open the screw on right side to open side cover and slide it
out).
3.
Notice a small connector (POLARITY connector) which is to be pulled out and plugged in again in the reverse way.
4.
Close the side cover and now the unit is ready for operation with POLARITY being
reversed.
UNPACKING
The High Voltage Unit, Type : HV501 has been thoroughly tested and is despatched in ready to
operate condition. However, on unpacking and prior to operation, it is advisable to check
visually and make sure that there is no visible damage caused in transist.
If any damage to the instrument be observed, do not switch ON the unit and report the matter
immediately to :
Customer Support Division
Nucleonix Systems Private Limited
Plot No : 162 A & B, PHASE II,
I.D.A.Cherlapally, Hyderabad - 500 051.
Ph: 91-040-27263701/32918055, FAX : 27262146
e-mail : [email protected] www.nucleonix.com
In all correspondance regarding the instrument, please mention the serial numbers of the unit,
month and year of purchase of the unit.
3
CHAPTER - I
INTRODUCTION
NUCLEONIX offers a standard High Voltage Unit Type : HV501 which is a compact STANDARD/
NIM Module (2bit). This is designed to meet the full HV bias requirement for a broad range of
Nuclear radiation detectors like G.M.detectors, ionization chambers, scintillation detectors, photo
multiplier tubes, etc,. Apart from this range of applications it also can be used as a high voltage
source in areas where high degree of regulation, stability and accuracy are required.Output can
be varied by a precise ten turn helipot and output indication is provided on a DPM.
FRONT VIEW
REAR VIEW
CHAPTER - II
SPECIFICATIONS
OUTPUT VOLTAGE
:
0V to 1500V continuously variable by a ten turn
helipot provided on front panel. (All specifications
are valid from 10V to 1500V).
OUTPUT INDICATION
:
Provided on a 3 1/2 digit LED, DPM.
OUTPUT CONNECTOR
:
UHF Socket (by default)
MHV Socket (if specified)
OUTPUT CURRENT
:
It can deliver upto a maximum of 1mA.
OUTPUT POLARITY
:
The unit is offered with a choice of both POSITIVE or
NEGATIVE POLARITY selectable by reversing a
polarity connector. (Provided inside).
LINE & LOAD REGULATION
:
Better than 0.005% of full scale.
RIPPLE
:
Less than 20 mV peak to peak.
PROTECTIONS
:
Indefinite protection against overload and short circuit.
Under overload the unit doesn't meet the
specifications whereas on removal of short circuit it
will recover by itself.
POWER REQUIREMENT
:
Draws power from a standard BIN / NIM Power
Supply or MINIBIN power supply.
+24V at 250mA
+12V at 50mA
-12V at 50mA
LINEARITY
:
Better than +/- 0.2%
TEMPERATURE STABILITY
:
Better than 100 PPM.(Better temperature stability can
be offered at extra cost).
MECHANICAL DIMENSIONS
:
Standard 2 bit.
REAR MODULE CONNECTOR
:
Amphenol connector Type:26-159-24P-H (24 pin
type) or NIM standard connector as per AEC
specifications TID 20893 (Rev) Type : AMP 2041865
optional.
CHAPTER - III
INSTALLATION
The High Voltage Unit Type : HV 501 is a standard HV unit two bit offered in modular form. It
is highly stable, well regulated with low ripple, noise and serves as a detector bias unit for various
nuclear detectors. It is specially designed to be as constituent part of MINIM Gamma Ray
Spectrometer and is used for biasing a Scintillation detector.
A.
INSERTING THE MODULE INTO THE BIN
The Bin is fitted with guides (SS rods) at the top and bottom to help guide and position the
modules. To insert the module into the Bin the projected edges at the top and bottom right hand
side of the module should be positioned into the S.S.rod channel grooves and pushed right-in
until the 24 pin / 42 pin power connectors of the bin and module are fully engaged. The captive
screws on the module should then be tightened.
B. POWER TO WORK THE MODULE
The input power to work the module is fed through the connector at the rear side of the module.
This connector mates with one of the output connector sockets of MB403 when the module is
slid into the bin.
CHAPTER - IV
OPERATING INSTRUCTIONS
A.
OPERATIONAL CONTROLS
1.
"ON" SWITCH
:
This is a toggle switch which facilitates one, either to choose
the EHT or disable the EHT when not required.
2.
EHT OUTPUT
:
This is an amphenol UHF socket Type : 83-798-H or MHV /
SHV connector (optional) from where one can draw the EHT
output, to the required load.
3.
EHT ADJUST
:
The EHT output can be adjusted by this knob. This knob is
fitted on to the shaft of a ten turn helipot.
B.
HOW TO OPERATE THE UNIT
1.
Insert the HV501 module into the bin as explained under "INSTALLATION".
2.
Switch 'ON' the bin and HV501 module to generate EHT and set the knob to obtain the
required EHT.
3.
With a proper multimeter (High impedance) check the 'EHT OUT' and ensure that EHT
OUT is proper and disconnect the multimeter.
4.
In a typical "Gamma Ray Spectrometer" system, Connect one end of the EHT cable
(cable with two end UHF/MHV male connector) to the "EHT OUTPUT" socket and the
other end to scintillation head EHT socket.
5.
Switch ON the bin and the ON switch of HV501 to generate EHT and set the knob to
obtain the required EHT setting.
C.
BLOCK DIAGRAM DESCRIPTION
This High Voltage Circuit employed in this design essentially consists of an external
driver oscillator (555 chip ). This generates square wave oscilations in the range of 25
to 35KHz. +12V supply from the bin is provided to this oscillator. This output is phase
split and fed to the differential driver ( Power Amplifier ) consisting of two power transistors,
fed to the primary of the R.F. transformer. The center tap of this R.F. is fed with the control
voltage, which in turn controls the primary swing.
The stepped up R.F output is fed to a voltage doubler and HV filter.
A fraction of this output through a bleeder is taken and compared at the op-amp input
with a standard reference voltage, applied through a ten turn pot. The output of this opamp inturn generates corrective control voltage through a regulator chip for EHT
regulation.
CHAPTER - V
BLOCK DIAGRAM DESCRIPTION
This High Voltage Circuit employed in this design essentially consists of an external driver
oscillator (555 chip ). This generates square wave oscilations in the range of 25 to 35KHz.
+12V supply from the bin is provided to this oscillator. This output is phase split and fed to the
differential driver ( Power Amplifier ) consisting of two power transistors, fed to the primary of
the R.F. transformer. The center tap of this R.F. is fed with the control voltage, which in turn
controls the primary swing.
The stepped up R.F output is fed to a voltage doubler and HV filter.
A fraction of this output through a bleeder is taken and compared at the op-amp input with a
standard reference voltage, applied through a ten turn pot. The output of this op-amp inturn
generates corrective control voltage through a regulator chip for EHT regulation.
BLOCK DIAGRAM of HIGH VOLTAGE (HV501)
HV
OUTPUT
PUSH PULL
DRIVER
RF TRANS
FORMER
VOLTAGE
DOUBLER
HV
FILTER
>
CONTROL VOLTAGE
V Ref
D
P
M
>
555 OSC
15 KHz
>
+24V
>
VOLTAGE
REGULATOR
+
-
TEN TURN
POT
>
D
i
s
p
l
a
y
CHAPTER - VI
AVAILING OF MAINTENANCE/ CALIBRATION SERVICES AND
WARRANTY CLAUSE (with in India)
6.1
GENERAL
As per the warranty clause of the company, we provide one year warranty during which period we provide
free service at our works. Hence in case of any mal-function in our instruments, you are requested to send
the unit back to our works by RPP/COURIER/SPEED POST PARCEL/GATI/XPS/door delivery. We shall
arrange immediate rectification/replacement within two weeks from the date of receipt of the equipment at
our place. Please note that the equipment will be serviced at our works only.
The equipment is to be sent to:
The Servicing Department
NUCLEONIX SYSTEMS PRIVATE LIMITED
Plot No: 162 A & B, PHASE II, I.D.A. Cherlapally,
Hyderabad - 500 051Ph: 040-27263701/329145448/32918055
E-mail: [email protected] www.nucleonix.com
For all the Radiation monitoring equipment, detectors built-in or external probes will not have one-year
warranty, but only inspection warranty at the time of supply is provided. Since detectors will / may have
fragile glass construction, we do not provide warranty. In case of failure of these components, Nucleonix will
supply detector replacement at cost-cost price.
Note: In respect of all types of portable radiation monitors, it may be necessary to checkup and recalibrate
the equipment once a year at our works.
6.2
EQUIPMENT REPAIRS / SERVICING POLICY (WITH IN INDIA)
(a)
During Warrantee
The following procedure is to be followed by the customers with in India for availing services/ repairing
facility during warrantee period.
●
●
●
●
●
●
●
●
●
●
Equipments are to be sent to our works for availing free repair services during warrantee, after the
customer receives approval from the customer support division, by sending an e-mail.
For all equipments, costing less than 6.0 lakhs one year warrantee & free service is offered, when the
equipments are sent to our works only. For larger systems such as installed systems, networked
systems, specialized systems, costing more than 6.0 lakhs during one year warrantee, free service
is offered at site. Field service Engineer will be deputed subject to warrantee terms & conditions.
This does not include personal computer related problems, for which local computer service provider
of the PC vendor is to be contacted. Also for software related problems online support will be provided.
Software support doesn't include cleaning of virus problems etc.
When the equipments are sent to our works for warrantee services, they are to be properly packed
with adequate cushion to prevent any transportation damages. Nucleonix Systems is not responsible
for damages or loss during transportation.
Packing / Freight charge is to be borne by customer when he sends the equipment to our works.
However when we return after servicing packing will be Nucleonix responsibility & Freight charges
will be to your account. Only services are free.
Please indicate in your correspondence equipment model & serial number.
All the equipments are to be sent to our works only on door delivery basis.
For Door Delivery Transportation contact XPS/GATI cargo in your city / town or a reliable courier
service to pick the consignment from your place. For their nearest local address & phone no's look
into their websites. Transit insurance if the customer feels is necessary it is to be covered.
Nucleonix Systems will not receive the equipments sent by other modes of transportation, such as
Rail/Road.
After servicing, equipments will be sent back by same mode of transport such as XPS/GATI/COURIER/
RPP.
●
●
(b)
All types of Radiation detectors, glass ware, PMTs etc which are fragile are not covered in warrantee,
if the failure is due to physical damage, external or internal due to shock, dropping, miss-handling
etc. If the failure is due to a natural fault then only it is covered under warrantee for a limited period of
three months. However complete electronics is covered for 1 year warrantee.
You can also send the equipment personally to our works for repairs either during or after warrantee,
after fixing up with our service dept (Customer Support Division). If possible we may repair on same
day or your person can stay for a day or two & get it repaired & or calibrated.
After warrantee Services
●
●
●
●
●
●
●
●
●
On expiry of 1yr warrantee if you like to send the equipment (low cost less than 6.0 lakhs) for repairs
to our works, you may please observe the following procedure.
Send an e-mail with details mentioning that you agree to pay service charges which includes: Basic
service charges per unit / module in the range of Rs: 2500 to Rs : 10,000 depending on the sophistication
of the unit calibration charges ( if applicable for your equipment) + cost of components + packing
charges + Return Freight charges @ actual.
Once our customer support department responds & requests you to despatch the equipment to our
works for repairs, you may do so by following the steps given below.
Followed by this you can send the equipment straight away if it is within 5 yrs old. If the equipment
is beyond 5 yrs old, then also you can send it for repairs, however only after you receive confirmation
from Customer Support Division, that it is repairable & is not an obsolete model. If the design is
obsolete then customer support division (CSD) may give you 'buy back' offer to replace with new
model or upgrade it with electronic circuit boards & enclosure.
For all installed equipments costing above Rs: 6.0 lakhs which are larger in size & for which field
servicing only is recommended, you can obtain a quotation with relevant details by sending an e-mail
& avail the services accordingly.
For all field servicing jobs, since we need to depute engineers, it is likely, to take time & also it will
cost more which includes Engineer's TA & DA etc., apart from basic service charges + cost of spares
etc. Please note that basic service charges will be different for different products depending upon
sophistication.
Also in some cases it may not be possible to fix-up the problems in the field itself, in such cases we
may advise you to send them to our works.
For all jobs to be serviced in the field, customer is requested to provide adequate details on the
nature of problems, to enable our engineer to come prepared with adequate spares.
For any additional information send an e-mail to [email protected], Atten: Customer support division.
6.3
EQUIPMENT REPAIRS / SERVICING POLICY (FOR EXPORTS)
Equipments, manufactured & exported are subjected to a well defined quality assurance (QA) plan &
Factory acceptance tests (FAT). Nucleonix systems has the following policy to provide maintenance support
to overseas customers either directly or through international dealers / distributors.
(a)
During & after warranty:
●
For minor problems, which can be handled by customers, servicing tips have been provided in the
user manual / servicing manual.
●
Also most of the equipments have built-in fault diagnostic features which will indicate to the user
nature of problem in the equipment. Based on the visual indication in the instrument Display, user
can take corrective action or contact Nucleonix systems by email for help.
●
Nucleonix systems will guide in localizing the defective part / module or sub-system by interacting
with the customer if required. Skype will be used for communication.
●
During warranty free replacement of sub-system or board (PCB) will be done. However customer has
to send defective sub-system back to Nucleonix system with-in 15 days on arranging replacement.
●
During & after warranty, any Freight charges & customs clearance charges are to be borne by
customers, both ways.
●
If it is a manufacturing defect, then Nucleonix system will bear the replacement cost of sub-system
/ unit. However any Freight charges & customs clearance charges in their country are to be borne by
customer.
●
After warranty, services will be similar to that of services during warranty. However, customer will
have to pay for cost of parts replaced, freight charges both ways & customs clearance charges in
both the countries. Nucleonix systems plans to introduce audio visuals on web or on CDs to facilitate
product demonstration, installation & minor maintenance very soon.
6.4
HOW TO AVAIL CALIBRATION SERVICES (FOR INDIAN CUSTOMERS)
Nucleonix Systems offers radiation calibration services to its customers. Calibration services are provided
for Nucleonix Systems manufactured products only, in general, as a company policy.
How to avail calibration services:
It is best advised that each of the Radiation monitors including Area monitors are calibrated once in a year.
When you want to send your Radiation monitor / Area monitor / Contamination monitor for calibration to our
works. You may send the equipment for calibration, by following the steps given below:
1.
2.
3.
4.
5.
6.
7.
6.5
Our standard calibration charges per equipment (All types of Radiation monitors including portable
survey meters, contamination monitors & Area Gamma Monitors) are Rs: 2500 + Packing + Freight
charges. You can email a 'work order' accepting these charges.
Email your work order and despatch / send the equipment to our works if it is 5 years old or less
including details of mode of transport sent with docket particulars.
Also mention in your work order & clearly indicate that you will agree to pay calibration charges &
also equipment repair charges additionally if the unit is faulty & requires repairs before one can take
it up for calibration.
You are requested to ensure good packing to avoid any transportation damages. Especially if there
are external detector probes, they are to be packed with sufficient soft foam to ensure no damage in
transportation.
Use only the specified following mode of transportation system for dispatching on door delivery
basis. XPS/GATI cargo / Courier/RPP/Speed Post parcel etc. Send the equipment on freight paid
basis. (Equipments sent by other methods such as Rail/Road etc will not be collected). Also you
can cover for transit insurance both ways if you wish. Nucleonix system is not responsible for any
transportation damages or loss during transportation both ways.
Immediately on receipt of the equipment, we will send an acknowledgement & also a proforma bill by
email/ post.
Based on the proforma bill, once we receive the payment, equipment will be dispatched back by
similar mode of transportation as mentioned above.
HOW TO AVAIL CALIBRATION SERVICES (FOR FOREIGN CUSTOMERS)
Foreign customers can calibrate Nucleonix make Radiation monitors/equipments in their country at any of
their accredited Radiation calibration labs. Nucleonix systems will be happy to provide any help and guidance
if needed, for calibration. Alternatively if you send the equipment here to India we can also provide calibration
services.
Calibration Standards Lab & Facility:
We have two calibration labs.
i.
ii.
Low Level Calibration Lab.
High Dose Rate Calibration lab.
Low Level Calibration Lab: This has a Cs-137, 165 mci standard. "Gamma Survey Instruments Calibrator"
from Amersham.
This calibration service has NIST Traceability standard. Calibration of all portable radiation monitors, survey
meters, contamination monitors, Area monitors etc., is carried out in this lab upto 1 R/hr max dose rates.
Gamma Survey instruments calibrator has Cs-137 source 161.5 mCi as on 05 Aug 2002. It is basically a
gamma survey instruments calibrator procured from AEA Technologies UK/USA. Has NIST traceability
accuracy within +/- 7%
Gamma Survey instruments calibrator has Cs-137 source 161.5 mCi as on 05 Aug 2002. It is basically a
gamma survey instruments calibrator procured from AEA Technologies UK/USA. Has NIST traceability
accuracy within +/- 7%
CRC-2 camera has Co-60 standard obtained from Bhabha Atomic Research Centre, Mumbai. It is a certified
source.
6.5
ANNUAL MAINTENANCE CONTRACT (AMC)
Annual maintenance contract (AMC) services:
For all sophisticated instruments & systems and also for installed monitors & networked systems in a
nuclear facility or a Radiological lab or in a Medical cyclotron facility where no. of instruments are networked,
it is advised that customer enters into an economical Annual maintenance contract with Nucleonix system.
Detailed AMC proposal can be obtained from our customer support division (CSD), by giving required
inputs.
Inputs required by our CSD to send you AMC proposal:
●
Name, year & data of purchase, Sl. Nos. of equipments, Model No's, No. of equipments for which
AMC is required. Additionally no. of calls per annum required for preventive & breakdown maintenance
may also be indicated.
Advantage of entering into AMC:
●
Equipment services offered will be prompt & timely
●
Nucleonix systems maintain required spares, spare tested PCBs, detectors & other critical
components which may become obsolete.
●
Obsolescence in electrons is quite rapid. If you enter into AMC guaranteed service for the period of
AMC will be the responsibility of Nucleonix Systems.
●
Nucleonix Systems will maintain Engineers at your disposal to attend to AMC calls on time
●
Without AMC prompt service calls are not guaranteed.
●
If some critical components become obsolete, then Nucleonix systems may request you to upgrade
the product with new model or new electronics which may be expensive if you are not under AMC.
Training on maintenance / servicing:
●
To a limited extent, we offer training on maintenance / repairs at our works to customers on chargeable
basis. Details can be obtained from our customer support division, by customers who may require
such services.