Patient Monitor User Manual
CO Measuring (Optional)
CO INIT ERR 4
CO INIT ERR 5
CO module
failure
CO INIT ERR 6
High
Stop using measuring function
of
CO
module,
notify
biomedical
engineer
or
Manufacturer’s service staff.
High
Stop using measuring function
of
CO
module,
notify
biomedical
engineer
or
Manufacturer’s service staff.
High
Stop using measuring function
of
CO
module,
notify
biomedical
engineer
or
Manufacturer’s service staff.
CO INIT ERR 7
CO INIT ERR 8
CO COMM STOP
CO module
failure or
communication
failure
CO COMM ERR
CO module
failure or
communication
failure
Prompt message (general alerts):
Message
Cause
CO TEMP EXCEED
Alarm Level
TB measuring value is beyond
High
measuring range.
CO MEASURE NEED
CO measuring needs parameters
PARAMENT
High
HEMOD
CALCULATE
PARAMENT
High
NEED
HEMOD
parameters
calculation
needs
CO TB SENSOR OFF
TB measuring cable falls off the
Low
monitor
CO TI SENSOR OFF
Sensor or cables fall off the
Low
monitor
18.6 Maintenance and Cleaning
WARNING
Before cleaning the monitor or the transducer, make sure that the equipment is switched
off and disconnected from the power line.
- 191 -