Download Vibrating Sample Magnetometer - Materials Research Facilities
Transcript
Section 5.5 VSM Electronics 5.4 Chapter 5 VSM Hardware Sample-Mounting Station The sample-mounting station (4096-110) is used to precisely locate and measure (within 0.5 mm) the sample offset, which is the distance between the sample and the bottom of the sample holder (see Figure 5-8). Figure 5-8. Sample-mounting station (4096-110) The VSM measurement algorithm requires an accurate measurement of this distance so that it can perform touchdown operations (See Chapters 1 and 4 for more details about touchdown operations). 5.5 VSM Electronics 5.5.1 Preamp Cable Assembly Figure 5-9 shows the preamp cable assembly (3096-300), which is the electrical connection between the coilset puck and the “PREAMP” port (JB-3) on the Model CM-B detection module. Integrated into the cable are dual 40-gain preamplifiers for up to two independent detection coils. The preamp end of the cable is plugged into the gray connector at the PPMS probe head. Figure 5-9. Preamp cable assembly (3096-300) 5-8 VSM Option User’s Manual, 1096-100, Rev. B0 February 2011 Quantum Design