4 Preparing the Stabil-Ion System for Operation
8.
Press CANCEL to return to the serial number display.
Press CANCEL again to return to the INFO Menu.
(Press CANCEL yet again to return to the status display for the gauge tube that
is turned on.)
NOTE: Be sure to complete data swapping before turning off the Controller.
The IGTMP area is cleared when the power is turned off.
Serial
IG No.’s
Info
Menu
4.18.2 Convectron Gauge Serial Number Display and Memory Data Swapping
From the factory, the Convectron memory will have two average memory data curves programmed.
The serial number of the gauge, and the CG channel or gas type to which it is assigned, is displayed
using the Serial # function under the INFO Menu.
1.
Press the ▼ arrow key until you see the information menu.
2.
Press the SELECT key to see the serial number display.
3.
Press the ▼ key again to see the CGA (a) gauge serial number.
4.
Press the SELECT key to select the CG data area.
5.
Press the ▼ arrow key to view the desired serial number display. Press ▼
repeatedly to view the successive displays (e.g. CGA a, or CGB b,…).
Info
Menu
Serial
IG No.’s
Serial
CG No.’s
CGA 00
#####
CGB 00
#####
If you don’t need to swap data, or if both memory curves are averaged data, skip to step 13.
If you want to swap the channels, CGA and CGB, for the serialized tube data, there is a data
swapping area called CG TMP where data can be exchanged. The numbers in the display
correspond to the Serial Numbers on the Convectron Gauge tubes. The way to change data from
CGB to TMP is as follows:
6.
Press the ▼ arrow key to display the swap information.
CG TMP CG
#####
take note of Serial No.
7.
Press the ▼ arrow key again to display the swap message.
8.
Press the SELECT key to see the swap action.
9.
Use the ▲ or ▼ arrow keys to select the desired gauge to swap data with.
Swap ???
Be Sure!
Swap TMP
& CGA
Swap TMP
& CGB
4-43
Series 370 Stabil-Ion Instruction Manual - 370119 - Rev. 05