Download Welch Allyn AED 20 Brochure

Transcript
55
5
Maintenance
This section contains information on inspecting, maintaining, cleaning, and servicing the
defibrillator.
Inspection
To ensure the readiness and optimum working condition of the defibrillator, it should be
inspected and tested daily, weekly, or monthly, depending on the frequency of its use.
The checks outlined in the operator checklists should be planned according to the
inspection schedule outlined below. The goal is to maintain the defibrillator in an
operation-ready state.
Also check new pads or other accessories for compatibility with the defibrillator when
they are first received.
Scheduling Inspections
Inspect each defibrillator regularly to ensure that it is ready for service when needed. The
following table presents guidelines for determining an appropriate inspection schedule for
your defibrillators.
Frequency of Use
Inspection Schedule
Weekly
Daily
Monthly
Weekly
Rarely
Monthly
Note
When powered by a non-rechargeable battery, the defibrillator must be used very
infrequently to prolong battery life.
Note
If the defibrillator is used more than once per month, it is recommended that
authorized service personnel perform a periodic inspection servicing at least once
per year.
Power-Up and Self-Test
Press the ON button to power-up the defibrillator and automatically perform a self-test. At
power-up, the following tests are performed: battery, main processor, memory and
program, stuck key, ECG preamp, and defibrillator.
If the built-in sensors in the defibrillator detect a problem prior to or during operation, the
defibrillator provides a voice or screen prompt indicating the problem.