GE MEDICAL SYSTEMS DIRECTION 2300000, REVISION 0 8-7-3 LOGIQ™ 5 SERVICE MANUAL BackPlane (FRU No. 502) (cont’d) 4.) From the left view, disconnect eight connectors (1-8) from the Backplane Assy. Refer to Figure 8117 on page 8-67 .. (2) (1) (3) (3) Back Plane (4) (5)(6)(7) (8) Figure 8-117 Disconnect 8 connectors from the left 5.) From the right view, disconnect two connectors (9-10). (11) (9) (12) (10) (13) (14) Figure 8-118 Disconnect 2 connectors from the right 6.) Unscrew four (4) screws (11-14) to remove the Backplane Assy. 7.) Pull the Backplane assy out carefully. WARNING 8-7-3-5 The weight of the Backplane with PCBs inside is approximately 20 kg. Two person is needed in the next step. Mounting procedure 1.) Install the new parts in the reverse order of removal. Chapter 8 - Replacement Procedures 8 - 67