Download Model DC100 Communication Interface Instruction
Transcript
6.5 Other Settings Correcting the calibration value manually (DCV/TC/ DI input module) Setting Correcting the calibration value manually (mA input module) Setting IM DC100-11E XZp1, p2, p3, ..... , p25<terminator> p1 Unit number (0 to 5) p2 Slot number (0 to 5) p3 DISPLAY p4 Zero calibration value for 20-mV range p5 Span calibration value for 20-mV range p6 Zero calibration value for 60-mV range p7 Span calibration value for 60-mV range p8 Zero calibration value for 200-mV range p9 Span calibration value for 200-mV range p10 Zero calibration value for 2-V range p11 Span calibration value for 2-V range p12 Zero calibration value for 6-V range p13 Span calibration value for 6-V range p14 Zero calibration value for 20-V range p15 Span calibration value for 20-V range p16 Zero calibration value for 50-V range p17 Span calibration value for 50-V range p18 Zero calibration value for Pt:1 mA p19 Span calibration value for Pt:1 mA p20 Zero calibration value for Pt:2 mA p21 Span calibration value for Pt:2 mA p22 Span calibration value for Pt:1 mA-H p23 Span calibration value for Pt:2 mA-H p24 Zero calibration value for Cu:2 mA p25 Span calibration value for Cu:2 mA XZp1, p2, p3, p4, p5<terminator> p1 Unit number (0 to 5) p2 Slot number (0 to 5) p3 DISPLAY p4 Zero calibration value for 20-mA range p5 Span calibration value for 20-mA range Correcting the calibration value manually (Strain input module) Setting Correcting the calibration value manually (universal module) Setting XZp1, p2.....p17<terminator> p1 Unit number (0 to 5) p2 Slot number (0 to 5) p3 DISPLAY p4 Zero calibration value for 20-mV range p5 Span calibration value for 20-mV range p6 Zero calibration value for 60-mV range p7 Span calibration value for 60-mV range p8 Zero calibration value for 200-mV range p9 Span calibration value for 200-mV range p10 Zero calibration value for 2-V range p11 Span calibration value for 2-V range p12 Zero calibration value for 6-V range p13 Span calibration value for 6-V range p14 Zero calibration value for 20-V range p15 Span calibration value for 20-V range p16 Zero calibration value for 50-V range p17 Span calibration value for 50-V range Example XZp1, p2, p3, ....., p9<terminator> p1 Unit number (0 to 5) p2 Slot number (0 to 5) p3 DISPLAY p4 Zero calibration value for 2k range p5 Span calibration value for 2k range p6 Zero calibration value for 20k range p7 Span calibration value for 20k range p8 Zero calibration value for 200k range p9 Span calibration value for 200k range Modify the span calibration value for the 60-mV range of the module in slot 1 of subunit 0 to 32000. XZ0, 1, DISPLAY, , , , 32000 Correcting the calibration value maually (Digital input module) Setting Example XZp1, p2, p3, p4, p5, p6, p7<terminator> p1 Unit number (0 to 5) p2 Slot number (0 to 5) p3 DISPLAY p4 Zero calibration value for 60mV range p5 Span calibration value for 60mV range p6 Zero calibration value for 6V range p7 Span calibration value for 6V range Modify the span calibration value for the 6V range of the module in slot 1 of subunit 0 to 32000. XZ0, 1, DISPLAY, , , , 32000 6-15 6 Commands 20 mV, 60 mV, 200 mV, 2 V, 6 V, 20 V, 50 V For a mA-input module: 20 mA For a strain input module: 2k_ZERO, 2k_SPAN, 20k_ZERO, 20k_SPAN, 200k_ZERO, 200k_SPAN, INIT For a digital-input module: 60mV, 6V Example Calibrate the 20-mV range on the module in slot 1 of subunit 0. XZ0, 1, CAL/EXEC, 20mV Comments • For DCV/TC/DI input modules, it is not possible to calibrate any RTD. • When calibrating the DC voltage range, apply the rated value of the voltage being calibrated to channel 3 of the module and short-circuit channel 2. • When calibrating an RTD, apply a 100-Ω input to channel 5 and short-circuit channel 4. • When calibrating a highly sensitive RTD, first calibrate it at the 60-mV and 200-mV ranges. This strategy enables the RTD to be calibrated more precisely. • When calibrating a mA-input module, apply the 20mA range being calibrated to channel 3 of the module and short-circuit channel 2. • If an INIT option is applied to a strain input module, the DC100 initializes all the preadjusted setpoints of the module (including “unopened” setpoints). After the use of the INIT option, always readjust all of the ranges (zero/span adjustment of the ‘2k,’ ’20k’ and ‘200k’ ranges). • Always calibrate a strain input module in the order of zero and span adjustments.