Download Model DC100 Communication Interface Instruction
Transcript
6.7 Data Output Request Command Comments • Since it takes an indefinite time to execute the DS command, return an ACK after processing. On the controller side, execute the following processing after receiving an ACK after transmitting the DS command. The format of the ACK status is as shown below. E0Cr+Lf The received command was normally processed. E1Cr+Lf There is an error in the received command. MI Initializes the built-in RAM. Mode Setting Operation mode MIp1<terminator> p1 YES(fixed) Example MIYES Comments • Initializing the built-in RAM clears all the data in the built-in RAM. Copy necessary files to the floppy disk/SCSI device. 6.7 Data Output Request Command TS Selects the output data. Mode All modes (A/D calibration data output is limited to A/D calibration mode only.) Setting TSp1<terminator> p1 0 Measured data output 1 Setting data output 2 Unit data output 3 RAM disk data output 4 Report data output 5 System configuration data output 8 A/D calibration data output 9 Setting data output in setup mode Comments • The setting for p1=0, 1, 2, 3 or 4 is effective only in the operation mode. • The setting for p1=8 is effective only in the A/D calibration mode. • The setting, p1=4 is effective when the instrument has the report function and one of hourly, daily and monthly reports is ON. FM Mode Setting Operation mode FMp1, p2, p3<terminator> p1 0 Outputs measured data in ASCII format. 1 Outputs measured data in binary format. 2 Outputs computed data in ASCII format. 3 Outputs computed data in binary format. p2 First output channel (001 to 560) First output computed channel (A01 to A60) p3 Last output channel (001 to 560) Last output computed channel (A01 to A60) Comments • You must specify data to be output with the TS command and execute “GET” or “ESC T” before sending the FM command. • The command selects the output format of measured/ computed data, either ASCII or binary, and the output channel. • If no input channel is recognized by the DC100 among the specified channels, a syntax error occurs. • The setting “p1=2” or “p1=3” is valid only for recorders with the optional computing function or installed a pulse module. • The optional computation channel number for a stand-alone model is A01 to A30. MF Mode Setting IM DC100-11E Sets the output formats of the measure ddata/computed file or periodic file data on a RAM disk. Operation mode MFp1, p2, p3, p4, p5, p6<terminator> p1 Output contents 0 File directory 1 Measured/computed file or periodic file data output in ASCII format. 2 Measured/computed file or periodic file data output in binary format 3 On/off information on channel on which specified file exists 4 Specified report data output in binary format p2 File name (up to eight characters) p3 First output channel number p4 Last output channel number p5 First output data number p6 Last output data number 6-19 6 Commands Selects the output format for measured/ computed data.