Download Model DC100 Communication Interface Instruction

Transcript
6.7 Data Output Request Command
Comments • Since it takes an indefinite time to execute the DS
command, return an ACK after processing. On the
controller side, execute the following processing
after receiving an ACK after transmitting the DS
command. The format of the ACK status is as shown
below.
E0Cr+Lf The received command was normally
processed.
E1Cr+Lf There is an error in the received
command.
MI
Initializes the built-in RAM.
Mode
Setting
Operation mode
MIp1<terminator>
p1
YES(fixed)
Example MIYES
Comments • Initializing the built-in RAM clears all the data in
the built-in RAM. Copy necessary files to the floppy
disk/SCSI device.
6.7 Data Output Request
Command
TS
Selects the output data.
Mode
All modes (A/D calibration data output is limited to A/D
calibration mode only.)
Setting
TSp1<terminator>
p1
0 Measured data output
1 Setting data output
2 Unit data output
3 RAM disk data output
4 Report data output
5 System configuration data output
8 A/D calibration data output
9 Setting data output in setup mode
Comments • The setting for p1=0, 1, 2, 3 or 4 is effective only in
the operation mode.
• The setting for p1=8 is effective only in the A/D
calibration mode.
• The setting, p1=4 is effective when the instrument
has the report function and one of hourly, daily and
monthly reports is ON.
FM
Mode
Setting
Operation mode
FMp1, p2, p3<terminator>
p1
0 Outputs measured data in ASCII format.
1 Outputs measured data in binary format.
2 Outputs computed data in ASCII format.
3 Outputs computed data in binary format.
p2
First output channel (001 to 560)
First output computed channel (A01 to A60)
p3
Last output channel (001 to 560)
Last output computed channel (A01 to A60)
Comments • You must specify data to be output with the TS
command and execute “GET” or “ESC T” before
sending the FM command.
• The command selects the output format of measured/
computed data, either ASCII or binary, and the
output channel.
• If no input channel is recognized by the DC100
among the specified channels, a syntax error occurs.
• The setting “p1=2” or “p1=3” is valid only for
recorders with the optional computing function or
installed a pulse module.
• The optional computation channel number for a
stand-alone model is A01 to A30.
MF
Mode
Setting
IM DC100-11E
Sets the output formats of the measure
ddata/computed file or periodic file data
on a RAM disk.
Operation mode
MFp1, p2, p3, p4, p5, p6<terminator>
p1
Output contents
0 File directory
1 Measured/computed file or periodic file
data output in ASCII format.
2 Measured/computed file or periodic file
data output in binary format
3 On/off information on channel on which
specified file exists
4 Specified report data output in binary
format
p2
File name (up to eight characters)
p3
First output channel number
p4
Last output channel number
p5
First output data number
p6
Last output data number
6-19
6
Commands
Selects the output format for measured/
computed data.