6.5 Other Settings
Correcting the calibration value manually (DCV/TC/
DI input module)
Setting
Correcting the calibration value manually (mA input
module)
Setting
IM DC100-11E
XZp1, p2, p3, ..... , p25<terminator>
p1
Unit number (0 to 5)
p2
Slot number (0 to 5)
p3
DISPLAY
p4
Zero calibration value for 20-mV range
p5
Span calibration value for 20-mV range
p6
Zero calibration value for 60-mV range
p7
Span calibration value for 60-mV range
p8
Zero calibration value for 200-mV range
p9
Span calibration value for 200-mV range
p10 Zero calibration value for 2-V range
p11 Span calibration value for 2-V range
p12 Zero calibration value for 6-V range
p13 Span calibration value for 6-V range
p14 Zero calibration value for 20-V range
p15 Span calibration value for 20-V range
p16 Zero calibration value for 50-V range
p17 Span calibration value for 50-V range
p18 Zero calibration value for Pt:1 mA
p19 Span calibration value for Pt:1 mA
p20 Zero calibration value for Pt:2 mA
p21 Span calibration value for Pt:2 mA
p22 Span calibration value for Pt:1 mA-H
p23 Span calibration value for Pt:2 mA-H
p24 Zero calibration value for Cu:2 mA
p25 Span calibration value for Cu:2 mA
XZp1, p2, p3, p4, p5<terminator>
p1
Unit number (0 to 5)
p2
Slot number (0 to 5)
p3
DISPLAY
p4
Zero calibration value for 20-mA range
p5
Span calibration value for 20-mA range
Correcting the calibration value manually (Strain
input module)
Setting
Correcting the calibration value manually (universal
module)
Setting
XZp1, p2.....p17<terminator>
p1
Unit number (0 to 5)
p2
Slot number (0 to 5)
p3
DISPLAY
p4
Zero calibration value for 20-mV range
p5
Span calibration value for 20-mV range
p6
Zero calibration value for 60-mV range
p7
Span calibration value for 60-mV range
p8
Zero calibration value for 200-mV range
p9
Span calibration value for 200-mV range
p10 Zero calibration value for 2-V range
p11 Span calibration value for 2-V range
p12 Zero calibration value for 6-V range
p13 Span calibration value for 6-V range
p14 Zero calibration value for 20-V range
p15 Span calibration value for 20-V range
p16 Zero calibration value for 50-V range
p17 Span calibration value for 50-V range
Example
XZp1, p2, p3, ....., p9<terminator>
p1
Unit number (0 to 5)
p2
Slot number (0 to 5)
p3
DISPLAY
p4
Zero calibration value for 2k range
p5
Span calibration value for 2k range
p6
Zero calibration value for 20k range
p7
Span calibration value for 20k range
p8
Zero calibration value for 200k range
p9
Span calibration value for 200k range
Modify the span calibration value for the 60-mV range
of the module in slot 1 of subunit 0 to 32000.
XZ0, 1, DISPLAY, , , , 32000
Correcting the calibration value maually (Digital input
module)
Setting
Example
XZp1, p2, p3, p4, p5, p6, p7<terminator>
p1
Unit number (0 to 5)
p2
Slot number (0 to 5)
p3
DISPLAY
p4
Zero calibration value for 60mV range
p5
Span calibration value for 60mV range
p6
Zero calibration value for 6V range
p7
Span calibration value for 6V range
Modify the span calibration value for the 6V range of
the module in slot 1 of subunit 0 to 32000.
XZ0, 1, DISPLAY, , , , 32000
6-15
6
Commands
20 mV, 60 mV, 200 mV, 2 V, 6 V, 20 V, 50 V
For a mA-input module:
20 mA
For a strain input module:
2k_ZERO,
2k_SPAN,
20k_ZERO,
20k_SPAN, 200k_ZERO, 200k_SPAN, INIT
For a digital-input module:
60mV, 6V
Example Calibrate the 20-mV range on the module in slot 1 of
subunit 0.
XZ0, 1, CAL/EXEC, 20mV
Comments • For DCV/TC/DI input modules, it is not possible to
calibrate any RTD.
• When calibrating the DC voltage range, apply the
rated value of the voltage being calibrated to channel
3 of the module and short-circuit channel 2.
• When calibrating an RTD, apply a 100-Ω input to
channel 5 and short-circuit channel 4.
• When calibrating a highly sensitive RTD, first
calibrate it at the 60-mV and 200-mV ranges. This
strategy enables the RTD to be calibrated more
precisely.
• When calibrating a mA-input module, apply the 20mA range being calibrated to channel 3 of the module
and short-circuit channel 2.
• If an INIT option is applied to a strain input module,
the DC100 initializes all the preadjusted setpoints
of the module (including “unopened” setpoints).
After the use of the INIT option, always readjust all
of the ranges (zero/span adjustment of the ‘2k,’ ’20k’
and ‘200k’ ranges).
• Always calibrate a strain input module in the order
of zero and span adjustments.