4.
GETTING STARTED
Upon receiving the Model 465L please do the following:
1. Verify that there is no apparent shipping damage. (If damage has occurred please advise
shipper first, then Teledyne API.)
2. Allow at least a 1-inch clearance for proper ventilation.
3. Connect sample inlet line(s) to the SAMPLE INLET (Figure 4-2).
NOTE: Sample tubing made from an intert material such as Teflon should be used to
minimize sample degradation.
CAUTION
Connect the exhaust fitting (Figure 4-2) to a suitable vent outside the monitor
area.
4. Ensure that there are no communication devices connected until after start-up is
complete.
WARNING: ELECTRICAL SHOCK HAZARD
DISCONNECT POWER TO THE AC MAINS BEFORE MAKING OR
REMOVING ANY ELECTRICAL CONNECTIONS TO THE INSTRUMENT.
4.1.
Wiring Requirements for NEMA Enclosures
Use appropriate wiring rated for this type of equipment, ensuring that it meets local and national safety
and building requirements.
Ensure that overcurrent protection is used (a 5 A circuit breaker is recommended), and that it fulfills the
following requirements:
be located as near to the instrument as possible
quickly and easily accessible
clearly labeled as the disconnecting device for this instrument
4.2.
Wiring Instructions
The electrical connection must be hard-wired to the 3-pin connector on the mainboard (Figure
4-1) labeled J23. AC power connection to the monitor should be made with 12-14 AWG
stranded copper wire, connected to the monitor as follows:
07795B DCN6806
19