Download Copyright
Transcript
Ambient Conditions Figure 24: 8.6 Load capacity on the rear side of the multiple socket outlet on the flexible supply unit Waste Heat and Cooling NOTICE Instruments may sustain damage Never install the system close to air conditioners or other ventilation systems. Air from air-conditioning systems or other ventilation equipment should not blow directly on the microscope. Use a high-performance air-conditioning system We recommend a high-performance air-conditioning system for operating the system to prevent large temperature fluctuations in the room. 8.6.1 TCS SP8 with flexible supply unit Waste heat of the complete system, including the argon laser 3.2 kW Waste heat of the argon laser 1.6 kW For cooling, the argon laser must be connected to the provided argon laser fan. The following space is required: Standard fan about 30 x 30 cm (1' x 1') Noise-reduced laser fan 50 x 60 cm (1'8" x 2') In addition, it is possible to connect the argon laser exhaust to an in-house exhaust system: 48 Hose length 400 cm (13'1"), provided by default Diameter of adapter piece 15 cm (6") Air flow rate 400 m³/h (14,126 ft³/h)