DT Zero Leak Volume (enter value) cc.
Volume of the drink tube zero leak

DT Zero Leak Max (enter value) sccm
If leak measures above this value the test fails

EV Target Pressure in.H2O
The EV leak test starts at this pressure

EV Zero Leak Max (enter value) sccm
If leak measures above this value, the test fails

EV Blockage Press (enter value) in.H2O
If backpressure measures above this value, the test fails

Bladder Test Pressure (enter value) psi
The starting pressure the bladder is inflated to during the bladder self
check

Bladder Inflate Time (enter value) sec
The maximum time allowed for the bladder to inflate during the bladder
self check

Bladder Decay Pressure (enter value) psi
The maximum pressure the bladder is allowed to decay during the
bladder self check

Bladder Decay Time (enter value) sec
The time allowed for the bladder pressure to decay during the bladder
self check

Mask Min. Purge Time (enter value) sec
The minimum time required to purge the mask of aerosol concentration

Mask Max. Purge Time (enter value) sec
The minimum time required to purge the mask of aerosol concentration

Mask Test Presample Time (enter value) sec
The Mask PF test will delay for this may seconds before staring the
mask sample

Mask Min Sample Time (enter value) sec
The Mask PF test will sample the mask port for at least this many
seconds

Mask Max Sample Time (enter value) sec
If the calculated mask sample time exceeds this many seconds, a low
particle concentration warning will be displayed

Mask Test Flow (enter value) L/min
The Mask PF test will run at this flow

Mask Vacuum (enter value) in.H2O
The Mask PF test will run at this vacuum

Protection Factor (enter value)
The Mask PF must read at least this high to pass the self check using
Mask Bypass Adapter