Model LTC-21 Cryogenic Temperature Controller
SDLM
Format: SDLM <0 = Oneshot>, <1 = Cyclical>;
LTC-21 Response: None.
Description:
Sets the Datalog Acquisition mode selection.
Related Commands:
QDLC? SDLC SDLS SDLI SDLT
SDLS
Format: SDLS <4 to 500>;
LTC-21 Response: None.
Description:
Sets the Datalog buffer size.
Related Commands:
QDLC? SDLC SDLI SDLM SDLT
SDLT
Format: SDLI <0 = stop datalog, 1 = start datalog>;
LTC-21 Response: None.
Description:
Sets the Datalog Interval selection.
Related Commands:
QDLC? SDLC SDLS SDLM SDLT
SDUR
Format: SDUR <0 = 0.5 sec., 1 = 1 sec. , 2 = 2 sec, 3 = 4 sec. , 4 = 8 sec., 5 = 16 sec. >;
LTC-21 Response: None.
Description:
Sets the Input Filter Time Constant or Display Update Rate index.
Related Commands:
QDUR?