For instructions on preparing the robot for use in a protocol, see the Infinium Assay Lab
Setup and Procedures Guide.
Refer to the figure below throughout this protocol. Note that all of the plate barcodes face to
the right.
Figure 99 Placing Alignment Fixtures and MSA3 Plate onto Robot Bed
A
B
MSA3 Plate
Robot BeadChip Alignment Fixtures
Verify MSA3 and BeadChips for Hybridization (LIMS only)(Optional)
1
Scan the barcode of the MSA3 plate.
2
Scan the barcodes of all the BeadChips you plan to hybridize with the plate. You can
scan up to 24 BeadChips.
NOTE
Only scan BeadChips that have been accessioned into the system. The BeadChip
type must match the type associated with this batch in Illumina LIMS.
Illumina Infinium HTS Assay Protocol Guide
149
Hybridize to BeadChip (Post-Amp)
Prepare the Robot