Controlling the 1260 Infinity MDS The RI Detector Menus 4 The RI Parameters Menu Screen The RI Parameters Menu screen is used to access the RI detector parameters. To access the RI Parameters screen, select the Parameters sub menu option from the RI Detector Menu screen. The items available in the menu are as follows: Purge Time This parameter defines the time delay between the RI purge valve opening and closing during a purge action selected from the RI Detector Menu screen. The time range and default value are shown below. Range: 0 – 600 s Default: 30 s Temperature This parameter sets the temperature of the RI detector in increments of 1 °C. There is also the option to turn the heater off. The temperature range and default value are shown below: Range: 30 – 60 °C with an Off option Default: Off Polarity This parameter sets the polarity of the RI detector output. If the polarity is set to "- ve", then the detector signal is flipped, i.e. sample peaks with a negative output in "+ve" mode become positive outputs in "- ve" mode. The polarity range and default value are shown below. Range: +ve / -ve Default: +ve Agilent MDS User Manual 79