Download The ALICE Silicon Pixel Detector Control and Calibration Systems
Transcript
132 Detector Calibration automation would be very complex if the data should be retrieved from the permanent data storage, analyzed offline and then used to update the CDB. Furthermore, in many calibrations methods, the triggers are generated by the off-detector electronics hence the required trigger information needed for the event building is missing. Data ALICE DAQ Control Detector Algorithms (DA) a) Standalone Runs b) Physics Runs Monitor Machine DA b RO Electronics DAQ Network DAQ FXS DCS Network FERO CDB General Purpose Network FXS-CDB Connector Offline Shuttle (Pre-Processing) Commands & Status FED FED Servers Servers (x2) (x2) Local Data Collectors (LDCs) DA a Reference Data DB OCDB FERO Control Reference Data Displayer FECS Optical Ethernet VME DIM Figure 5.1: A DAQ ACTIVE calibration scenario block diagram. In order to cope with these specifications I foresaw that, during the SPD calibration, the produced calibration raw data are stored locally in the SPD LDCs and they are analyzed online at the end of each run. The event building is as well suppressed during the SPD calibration. The DAQ configuration swap between the physics data taking mode and the calibration mode is performed, if needed, automatically by the ECS at the beginning of each run. When the DAQ is ready for calibration data taking, the ECS sends to the SPD FSM top-node the start calibration command. Using the command information, the top-node decides if the calibration to be performed is ei-