Download here - Element14
Transcript
SCPI Command Reference 5-53 :ALPHa <NRf> [:SENSe[1]]:TEMPerature:FRTD:ALPHa <NRf> Specify alpha value Parameters <NRf> = 0 to 0.01 Query :ALPHa? Description This command is used to check and/or change the alpha value. Remember that if you change the present alpha value, the FRTD TYPE will change to USER. Specify FRTD alpha value Query value of ALPHa :BETA <NRf> [:SENSe[1]]:TEMPerature:FRTD:BETA <NRf> Specify beta value Parameters <NRf> = 0 to 1.00 Query :BETA? Description This command is used to check and/or change the beta value. Remember that if you change the present beta value, the FRTD TYPE will change to USER. Specify FRTD beta value Query value of BETA :DELTa <NRf> [:SENSe[1]]:TEMPerature:FRTD:DELTa <NRf> Specify delta value Parameters <NRf> = 0 to 5.00 Query :DELTa? Description This command is used to check and/or change the delta value. Remember that if you change the present delta value, the FRTD TYPE will change to USER. Specify FRTD delta value Query value of DELTa :DIODe command :RANGe[:UPPer] <NRf> [:SENSe[1]]:DIODe:CURRent:RANGe[:UPPer] <NRf> Select current range for diode test Parameters <NRf> = 0 to 1e-3 Specify diode test current Query [UPPer]? Description There are three current ranges available for the diode test: 10µA range, 100µA range, and the 1mA range. Range is selected by using this command to specify the expected current for the diode under test. The instrument will then automatically select the appropriate range. Query selected range