Download User`s Manual
Transcript
EVaRT 5.0 User’s Manual Appendix B: Analog Input Hardware and Software Which USB Device is Channels 1-32, which is 33-64? The first device you plug in should be channels 1-32 in your EVaRT software. The second device should be channels 33-64 and so on for more devices. If you are not sure: When you connect to the Cameras in EVaRT 5.0.3 or later, the channel numbers (1-32) and the Serial Number of the USB device are reported in the dialog box. The Serial Number for the USB-6218 device is located on the bottom of the USB device. Clock Wiring for Single and Multiple USB Devices Eagle and Hawk cameras require a 2.7 kΩ pull-up resistor from +5 Volts to the PFI-0 pin as shown Figure B-9. Clock Wiring for Mutiple Devices A Single Pull-up resistor with a 2.7 kΩ (or close) value will work for connecting up to 6 USB devices together as shown below. You need only to connect PFI0 and D GND signals in parallel for clocking the multiple USB Devices. A single pull-up resistor works for all devices. USB 2.0 Cable Lengths The USB-6218 comes with a one meter USB cable. It has been tested with longer USB cable lengths up to 16 ft. (5 meters) and it works well with either and any mix of cables. B-11