Chapter 9 – Errors
Error message
Possible causes
What you can do
Error 12: Door locking
Locking pin of the door is stiff
Check the smooth running of the door locking
pin
On repeated occurrence notify your specialist
dealer
see warning text Warning no feed water,
however this error appears after the start of a
program
Error 14: No feed water
Error 18: Sensor: ...Input:
…
The device-internal inspection
On repeated occurrence notify your specialist
of the sensors for temperature, dealer
pressure or conductivity
resulted in a too large deviation,
the message can occur when
the device is switched on or
during a program
Error 21: Preheating
The monitoring time from
switching on the pre-heating
until the achievement of the
respective pre-heating
temperature was exceeded
On repeated occurrence select
option Automatic pre-heating No ( see
page 23) and notify specialist dealers
Error 22: Overheat
Preheating
The maximum preheating
temperature was exceeded
On repeated occurrence select
option Automatic pre-heating No ( see
page 23, Select automatic pre-heating) and
notify specialist dealers
Error 23: Current
Monitoring time for the pressure Check flow filter lower front in the chamber floor
release in the outflow procedure
during fractionations was
exceeded
Flow filter soiled
On repeated occurrence notify your specialist
dealer
Error 26: A/D converting
The maximum admissible
On repeated occurrence notify your specialist
deviation of the computerdealer
internal signal conditioning (A/D
converting) was exceeded
Error 27: Temp.Sens. def
1.2
The maximum admissible
deviation between the two
temperature sensors for the
steam temperature was
exceeded
Error 31:System leak
During the program vacuum test Repeat vacuum test,
the permitted pressure
if renewed error message, notify specialist
maximum was exceeded (very dealers
large leak) after achieving the
evacuation pressure
Error 32: Power loss /
The loss of the operating
voltage occurred after the start
of a program
Sterilize sterile filter
On repeated occurrence notify your specialist
dealer
The error message occurs after the operating
voltage is again available
Check building-side installation, If no defects
are detectable, notify MELAG customer service
In the event of power failure after a program
50