Download Cyto Pulse Sciences, Inc.
Transcript
PA-101 User Manual: Rev.1-1/05 The Cyto Pulse cell fusion coaxial electrode has been optimized to convert the PA-101 sine wave voltage to an efficient non-uniform alternating field for cell alignment. It also converts the PA-4000 high voltage pulse(s) to a DC pulse electric field for cell fusion. The alignment process using dielectrophoresis is a function of: • The time that the AC wave is on • The intensity of the AC electric field • The frequency of the AC electric field • The non-uniformity of the AC electric field The first three parameters are selected using the PA-4000 interface software, and the last parameter is a property of the coaxial electrode The following chart shows the approximate intensity of the AC field for the three Cyto Pulse electrodes at a set ac voltage of 75 volts-peak. The field is non-uniform the values below are given at the inner electrode, mid way, and at the outer electrode. Part No. FE-25/400 FE-25/800 FE-20/1000 Gap mm 2.5 2.5 2.0 Inner V/cm 179 179 415 Mid V/cm 135 135 373 Outer V/cm 109 109 340 2.3.2 Step 2 – Fusion Pulse The second step in electrofusion is to apply one or more high voltage pulses to the cells, inducing membrane fusion. The voltage required must be above a threshold to induce membrane breakdown and below a maximum voltage causing cell death. Threshold voltage is approximately 0.5-1.5 volts across the cell membrane (1). The voltage across a cell is approximately equal to: Vt = 1.5 ∗ r ∗ E cos θ volts where: r = cell radius in cm E = electric field intensity in v/cm Θ = angle of the membrane in relation to the direction of the field Electric fields 2-3 times threshold can be used. This should be optimized for the cells used. Multiple fusion pulses may be more efficient than a single pulse. The table below gives the approximate electric field intensity for a pulse of 800 volt in the Cyto Pulse electrodes. Part No. FE-25/400 FE-25/800 FE-20/1000 2-6 Gap mm 2.5 2.5 2.0 Inner V/cm 4087 4087 4430 Mid V/cm 3107 3107 3990 Outer V/cm 2496 2496 3624 Cyto Pulse Sciences, Inc., P. O. Box 609, Columbia, MD 21045 USA 410-787-1890