Download PDA-100 Photodiode Array Detector Operator`s Manual
Transcript
4 • Troubleshooting 4.14 Error Message: Not Enough Light to Perform Operation • Light signal is not reaching the data processor Make sure that at least one lamp is turned on. • Low light throughput Remove the cell from the optical bench (see Section 5.4). Insert a business card into the cell mounting location and check that bright light is shining onto the card. If you can see bright light shining onto the business card, remove the card from the optical bench, run the lamp intensity calibration (see Section 3.6), and then autozero the detector. If no error message is displayed, flush the cell with eluent and/or clean the cell as instructed in (see Section 5.2). Reinstall the cell in the optical bench and autozero the detector again to verify that the problem has been resolved. If the error message appears again, contact Dionex Technical Support for assistance. If you see a weak light (or no light at all) shining onto the business card, the lamp intensity is too low. Check that the flange of the new lamp is aligned with the two locating pins in the optical bench (see Figure 5-3); if the flange is aligned, the lamp is defective and should be replaced. (For lamp replacement instructions, see Section 5.7 or Section 5.8 for the deuterium lamp or tungsten lamp, respectively.) If the error message appears again, contact Dionex Technical Support for assistance. • Blockage or low light transmission in cell Flush the cell with eluent. Doc. 031644-02 9/03 4-13