MC-133 / MC-133i Power calibrator
Example:
PHAR:VOLT1:MOD 12.45
Note:
Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:VOLT<n>:MOD:SHAP(?) <CPD> { SIN | RECT | OFF }
This command sets type of modulation for voltage output in channel <n>.
<CPD>
•
SIN – modulation using sinusoidal wave shape
•
RECT – modulation using rectangular wave shape
•
OFF - disables modulation
If query is sent, calibrator returns SIN, RECT or OFF.
Example:
PHAR:VOLT1:MOD:SHAP SIN
Note:
Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:VOLT<n>:MOD:DUTY(?) <DNPD>
This command sets the duty cycle of rectangular modulation for voltage output in channel <n>.
<DNPD>
Represents the duty cycle in %.
If query is sent, calibrator returns the duty cycle using standard exponential format. Example: 50%
is returned as 5.000000e+001.
Example:
PHAR:VOLT1:MOD:DUTY 50
Note:
Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:CURR<n>(?) <DNPD>
This command sets the output current value in channel <n>
<DNPD>
Parameter represents the value of the output current expressed in A.
<n>
Parameter represents the output channel (1, 2 or 3).
If query is sent, calibrator returns the value of output current using standard exponential format.
Example: 11.012 A is returned as 1.101200e+001.