Download Isco6700
Transcript
6700 SAMPLERS Chapter 5 Reference S AMPLE E VENT Chapter Contents • Sample Event • Sample Volume • Suction Head • Rinses and Retries • Pacing • Distribution • Sampler Enable • Pauses/Resumes • Start Times • Other Functions • Manual Functions • Programming Style • Hardware Setup • System IDs • Memory • Pressurized Lines • Command Driven Operation A sample event consists of a complete sampling cycle. A sample event includes presample and postsample purges, line rinses, sampling retries, and deliveries of sample volumes. Only one sample volume is placed in a given bottle during a sample event, however, depending on the distribution, multiple bottles may receive sample volumes. S AMPLE VOLUME The sample volume is the discrete, programmed amount of liquid delivered. During a sample event the sampler may deliver a sample volume to several bottles, depending on the distribution. “Sample volumes” and “samples,” are used interchangeably. Extended programming adds a flow-dependent sample volume option for uniform time-paced programs. If a flow module is attached, the flow source may be the module’s flow volume. Otherwise it is the flow pulse count received at the external flow meter connector. You enter the amount of flow required for each 10 ml of liquid. When the sample event is initiated, the sample volume will be based on the flow that has occurred since the last sample. This sample volume is will be at least 20 ml, but not more than the bottle volume (or 9990 ml, whichever is smaller). With flow-dependent sample volumes, the sampler resets the flow volume count at the start time. Therefore, the sampler will not take a sample at the start time. S UCTION H EAD Extended Screen for Suction Head Settings 1 AUTO SUCTION HEAD ENTER HEAD 2 SUCTION HEAD OF ___ ft (min - max) Suction head is the vertical distance between the surface of the liquid source and the sampler’s pump. At each sample event, the sampler determines the suction head automatically. In extended programming, you can manually enter the suction head if you want. Select ENTER HEAD only when the head remains stable and you can measure it accurately. 81