5.3
Measuring results
By repeatedly pressing the key RESULT, the following measuring results can
be indicated ("measuring result" here describes the result of a dynamic measurement, i.e. a number of individual measuring results):
- COMBINE current measuring result as per selected transducer combination
- MAX
maximum result during measurement (maximum value)
- MIN
minimum result during measurement (minimum value)
- (MX+M)/2 arithmetic mean value of maximum and minimum result
- (MAX-MIN) difference between maximum and minimum result
- MEAN
mean value of all individual measuring results assessed during
the measuring time Tm
Provided that the statistical evaluation was activated by the function "Stat. on",
the statistical results of a measuring series can be indicated as well:
-
n
xn
x (x/)
s
Xmax
Xmin
R
number of measurements or results
last measuring result
x-bar, mean value of all results of all measurements
standard deviation of these values from the mean value
maximum individual measuring result
minimum individual measuring result
range, difference between the largest and smallest result of all
measurements (Xmax - Xmin)
The measuring results to be indicated have to be selected beforehand in the
instrument configuration.
For all statistical evaluations it is necessary to beforehand set "Stat. on" in the
instrument configuration and select a measuring result, e.g. (MAX - MIN).
21