MI 3295 Step Contact Voltage Measuring System 10 Technical specifications Technical specifications 10.1 Step voltage, Contact voltage Measuring range Um 0.01 19.99 mV 20.0 199.9 mV 200 1999 mV 2.00 19.99 V 20.0 59.9 V Resolution 0.01 mV 0.1 mV 1 mV 0.01 V 0.1V Accuracy Calculated measuring range U 0.0 199.9 V 200 999 V Resolution Accuracy 0.1 V 1V calculated value* (2 % of reading + 2 dig) *Displayed Step / Contact voltage is obtained on base of calculation: US=Umeas ·Ifault/Igen; UC=Umeas ·Ifault/Igen; Ifault (selectable)....... 10 A ... 200 kA Input resistance (selectable): 1 kΩ, 1 MΩ Noise cancelation: DSP filtering 55 Hz, 64 dB rejection of 50 (60) Hz noise Test terminals: Test connector Meter 10.2 Current Measuring range 0.00 9.99 A 10.0 99.9 A Resolution 0.01 A 0.1 A Accuracy (3 % of reading + 5 dig) (3 % of reading + 3 dig) Current generator: 55 A max Test voltage: < 55 V Test frequency: 55 Hz Test terminals: C1/H - C2/E Station 10.3 Resistance to earth Measuring range 0.001 1.999 Ω 2.00 19.99 Ω 20.0 99.9 Ω 100.0 199.9 Ω Resolution 0.001 Ω 0.01 Ω 0.1 Ω 0.1 Ω Open circuit voltage ......................................... < 50 VAC Test current...............................................................< 7.5 A 40 Accuracy (2 % of reading + 5 dig) (5 % of reading)