depress CHG. This function also limits the Static Decay Meter ability to measure decay times longer than this time out. 8. When the CUTOFF threshold selected is reached, the unit will automatically switch to the TEST OVER state, the Sensor Head Shutter will close, and the Decay Time measurement, in seconds, will be displayed on the DECAY TIME readout. The unit will remain in the TEST OVER state until the ZERO/STBY or CHG button is depressed. 9. To repeat the test measurement, depress the CHG button and, when the SAMPLE CHARGE meter reads full scale, depress TEST. For the greatest accuracy,, it is recommended that the zero calibration point be checked between decay time measurements and the ZERO control adjusted if necessary. As noted earlier, calibration of the zero point is always done with the system in ZERO/STANDBY. NOTE: For additional information on sample testing, please refer to Section 4.0 "Testing Hints and Other Tests". 3.3 AUTO Mode In this mode, the Model 406D Static Decay Meter will conduct up to 9 measurement cycles automatically, including automatic zero calibration. This mode cannot be used with samples that have an initial or a residual charge. 3.3.1 Initial Set Up Follow the procedures outlined in Section 3.1. 3.3.2 Full Scale Calibration With the unit in MANUAL and ZERO/STANDBY, zero the SAMPLE CHARGE meter by adjusting the ZERO control. Select the CHARGING VOLTAGE polarity by depressing either the + or - button. Wait approximately 2 seconds, then set the CHARGING VOLTAGE meter to either + or - 5kV using the HIGH VOLTAGE ADJUST Control. Depress the CHG button and adjust the FULL SCALE calibrate control so that the SAMPLE CHARGE meter reads ±5kV. Unless this adjustment is properly made before the AUTO mode is selected, the Automatic Mode may not function properly. The FULL SCALE ADJUST knob should point approximately to the NORM position. For non-planar material, other settings will apply to achieve a full scale reading. 3.3.3 Sample Testing After the unit has been calibrated, select the AUTO mode by depressing the AUTO/MAN button. The button will latch in the down position, the SAMPLE CHARGE meter will read zero, the unit will return to the ZERO/STBY state and the AUTO mode readout will be illuminated and display zero. 15