Isco Flowlink 5® Software
Site Window - 4100/4200/6700
Reading type
Readings/Block
Rainfall
124
PH
124
Dissolved Oxygen
124
Level
62
Velocity
62
Temperature
62
Conductivity*
62
4-20 percent
62
Flow Rates
31
* Conductivity includes reading types reported as specific
conductance, salinity, total dissolved solids, and conductivity.
Calculating the partition size - apply the Time Span in Days,
Time Span in Hours, Reading Interval in minutes, and Readings/
block to the equation below.
Where:
D = Time Span Days
H = Time Span Hours
R = Reading Interval in minutes
B = Readings/block
P = Partition Size
The following example applies typical partition settings to the
equation.
Example: A level partition stores readings at 5 minute intervals
over a time span of 9 days, 1 hour.
Therefore, this partition uses 5376 bytes.
Calculating the size of a Sample Event Partition - Multiply the
number of samples by four to calculate the partition size in bytes.
Note: A sample event partition does not use reading intervals
and time spans. It simply stores sample events as they happen.
When you set up a sample event partition, set the “number of
samples” to the maximum number of expected samples. When
the instrument creates the partition, the number of samples is
rounded up to fill a complete block of memory. Each block of
memory holds 32 sample events.
73