5B SAMPLE AND REAGENT PROBES 5B1 Summary of Preventive Maintenance WARNING To prevent eye damage, wear safety glasses. Maintenance Material 1. Clean sample and reagent probes. 10-mL syringe 2. Check alignment of sample and reagent probes in reaction cuvettes. 0.030-inch gauge (P/N 192-R263-01) 3. Check alignment of sample probe in sample cups, and reagent probe in reagent boats. N/A 4. Check reagent probe heater. YSI 4502 Thermistor probe YSI 45 CU Thermometer Modified reaction tray cover Figure 5B-1 SAMPLE PROBE CLEANING 5B2 Figure 5B-2 REAGENT PROBE CLEANING Cleaning New Sample and Reagent Probes 1. Remove the sample probe assembly from the probe arm. 2. Connect a 10-mL syringe to the sample probe line as illustrated in Figure 5B-1. (Refer to Figure 5B-2 when cleaning reagent probes.) 3. Immerse the probe tip in a beaker filled with System Wash Solution (T01-1287). 4. Use the syringe to force wash solution back and forth through the probe for approximately five minutes. Occasionally remove the probe tip from the wash to aspirate air in 0.5-mL slugs (approx.). The air bubbles will help to scrub the inner walls. 5. Repeat Step 3 with distilled water. Rinse thoroughly. 5-2