Appendix C —Arrhythmia and ST Analysis
Setup Windows
ALARMS
Select Other Alarms Setups from the ALARMS menu to access the
two additional settings windows for ST and Arrhythmia. The
arrhythmia alarms window will not be available if arrhythmia is turned
OFF in the PARAMS menu.
Select Arrhythmia Setup to set the alarms for arrhythmia.
EXIT
Alarm Volume
ECG Lead Fail
Patient size
Heart Rate
SpO2
NIBP Systolic
NIBP Diastolic
NIBP Mean
Temperature
Respiration
CO2 Inspired
CO2 Expired
Apnea
5
MEDIUM
Adult
HIGH
150
OFF
200
100
150
100.0
36
1.5
7.5
60 seconds
°F
kPa
kPa
Other Alarm Setups
LOW
40
90
50
30
50
93.0
OFF
OFF
1.5
Arrhythmia Setup
Figure C-3: Select Arrhythmia Setup
Alarm priorities that appear in yellow are fully adjustable. Asystole,
VFIB, and VTACH are permanently set to high priority.
EXIT
Heart Rate
PVC Rate
PVC Run
UNITS
BPM
/min
Asystole
VFIB
VTACH
VT > 2
Bigeminy
Trigeminy
PVC Couplet
Irregular HR
Arrhythmia Relearn
HIGH
150
LOW
40
2
2
Min. HR
ALARM PRIORITY
HIGH
DISABLED
DISABLED
160
HIGH
HIGH
HIGH
MEDIUM
MESSAGE
MESSAGE
MESSAGE
MESSAGE
MESSAGE
<<< BACK
Figure C-4: Extended ALARMS Window for Arrhythmia
Page C-8
nGenuityTM 8100E Series Operator Manual
Criticare Systems, Inc.