Download User Manual
Transcript
CritiCool Pro™ Operating Manual Maintenance CAUTION! Do not use the steam autoclave method to sterilize reusable sensors and disposable sensor adapters. Cleaning and Disinfecting Procedures for Reusable Sensors Cleaning Probes should be cleaned with a mild detergent and water to remove excess bioburden and improve the effectiveness of disinfecting and sterilization. Disinfecting • For low-level cleaning use Cidex / glutaraldehyde • For high-level cleaning use Cidex / glutaraldehyde dilute bleach, 70% isopropyl alcohol Filter Replacement Note: NOTE An additional filter is supplied in the accessory box. The filter must be replaced every twelve months. 6 6-2 To replacement the filter (by the hospital personnel): 1. Drain the water tank (see Table 7-1, on page 7-2). 2. Remove the rear cover: 3. Unscrew the thumb captive screws at the bottom of the cover. 4. Pull the bottom part of the cover towards you and then down to release the lip from the chassis. 5. Unscrew the filter clamp screw and remove the filter clamp (see Figure 6-2). 6. Release both the water-in and water-out tubes from the filter assembly by pressing the release ring of each end of the filter and pulling the tubes from the filter. 7. Dispose of the old filter. MTRE®