2.8 Usage of Oxygen
• The presence of oxygen can speed up combustion of inflammable materials.
• If oxygen has been prescribed, connect the oxygen supply tube to the
appropriate oxygen port of the nasal mask or breathing system connector.
• At a fixed flow rate of supplemental oxygen flow, the inhaled oxygen concentration will vary, depending on the pressure delivered, patient’s breathing pattern, mask selection, and leak rate.
• When oxygen is used with the iSleep 10, the oxygen flow must be turned
off when the iSleep 10 is not operating. If the iSleep 10 is not in operation,
and the oxygen flow is left on, oxygen delivered into the patient tubing may
accumulate within the iSleep 10 enclosure. Oxygen accumulated in the
iSleep 10 enclosure will create a risk of fire.
• Ventilate the room adequately.
• Do not smoke in a room where oxygen is being used.
• Naked light bulbs and other sources of ignition must be kept a minimum of
2 metres away from the oxygen cylinder or any part of the patient circuit.
• Do not use aerosols or solvents close to the oxygen supply, even when the
oxygen supply is shut off.