Download 1 - MedWrench
Transcript
Preventive Maintenance Access and Replacement Instructions 5.6 Access and Replacement Instructions 5.6.1 Pump Maintenance Instructions Replacement Diaphragms P/N 0348-00-0129 Replacement Pressure Head P/N 0997-00-0402 Replacement Vacuum Head P/N 0997-00-0401 The pressure/vacuum pump is fitted with neoprene diaphragms which should be replaced at 2500 hour intervals to enhance System reliability. 1. (See FIGURE 5-1) Remove the System from the hospital cart. Remove the right side cover and the left lower cover from the System. Remove the two nuts at the rear of the pump compartment and remove the pressure and vacuum lines through the access hole provided. Remove the two screws at the base of the pump assembly and slide the pump housing out. Remove the 14 screws that retain the cover and remove the cover. FIGURE 5-1 Pump Assembly Removal CS300™ Service Manual 0070-10-0689 5-9