Download Temperature Verification System Overview 1
Transcript
To run the Temperature Non-Uniformity Test: Step Action 13 When the GeneAmp 2400 completes the Temperature Non-Uniformity Test, it displays the actual values gathered during testing. 14 Press F1-Accept or F5-Cancel. 15 Turn off the digital thermometer, remove the probe assembly from the sample block and clean the oil from the sample block. Temperature The maximum temperature measured in any well at the 94 °C setpoint minus the Non-Uniformity minimum temperature measured in any well at the 94 °C setpoint must be equal to or Specifications less than 1.0 °C. This can be expressed as: Max Temp 94 °C – Min Temp 94 °C < 1.0 °C The maximum temperature measured in any well at the 37 °C setpoint minus the minimum temperature measured in any well at the 37 °C setpoint must be equal to or less than 1.0 °C. This can be expressed as: Max Temp 37 °C – Min Temp 37 °C < 1.0 °C Temperature Verification Tests for the 2400 2-13