Troubleshooting Guide.
Problem
Cause
Solution
Light does not illuminate
No power to light
Light bulb has expired
Transformer is expired
Secondary windings on
transformers are expired
Main power switch is bad
Short or open in power line
form transformer.
Light head tension springs are
loose
Vertical arm tension spring
needs to be adjusted
Light post is not level
Check power
Replace light bulb
Replace transformer
Replace transformer
Only high or low function
works
Light turns on and off then
the light head is moved
Light head drifts down
Vertical arm drifts up or down
Light drifts from left to right
Replace main power switch
Check each power line for
short or open. Then replace.
Tighten light head spring nuts.
Adjust vertical arm tension.
Level light post.