LOGIQ 400 PROPRIETARY MANUAL GE MEDICAL SYSTEMS REV 0 2127662 2–4–1 Self Diagnosis (Continued) TEST NUMBER 13 TEST NAME TIME LINE MEMORY TEST PURPOSE To test the TLMP block of TLMC Assy or TDCB Assy. The BMRC Assy sends data via the USC bus to the memory on TLMP block. The DSP of TLMP block reads transmitted data from the memory. Results is sent to Main Computer. CRITERIA No Error : The test must successfully complete without error. DISPLAY FORMAT OF RESULTS CFM model: TLMC Passed or TLMC Failed B/W model: TDCB Passed or TDCB Failed TIME LINE MEMORY TEST MENU ILLUSTRATION 2–34 Action if the test fails Change TLMC Assy or TDCB Assy. The following Assemblies are necessary for this test CFM model: PS, MPU, BMRC, ASPR, DOPC, DSCC, TLMC, VIDO (or VIPB), KB, MON B/W model: PS, MPU, BMRC, ASPR, TDCB, VIDM, KB, MON 2–50 DIAGNOSTICS