Download Technical Publications

Transcript
LOGIQ 400 PROPRIETARY MANUAL
GE MEDICAL SYSTEMS
REV 0
2127662
2–4–1 Self Diagnosis (Continued)
TEST NUMBER 13
TEST NAME
TIME LINE MEMORY TEST
PURPOSE
To test the TLMP block of TLMC Assy or TDCB Assy. The BMRC Assy sends data via the USC bus to
the memory on TLMP block. The DSP of TLMP block reads transmitted data from the memory. Results is sent to Main Computer.
CRITERIA
No Error : The test must successfully complete without error.
DISPLAY FORMAT OF RESULTS
CFM model: TLMC Passed or TLMC Failed
B/W model: TDCB Passed or TDCB Failed
TIME LINE MEMORY TEST MENU
ILLUSTRATION 2–34
Action if the test fails
Change TLMC Assy or TDCB Assy.
The following Assemblies are necessary for this test
CFM model: PS, MPU, BMRC, ASPR, DOPC, DSCC, TLMC, VIDO (or VIPB), KB, MON
B/W model: PS, MPU, BMRC, ASPR, TDCB, VIDM, KB, MON
2–50
DIAGNOSTICS