Download Query Only
Transcript
Command descriptions
Syntax
Arguments
CALibration:CATalog?[{ALL|<subsystem>}[,{ALL|<area>}]]
ALL – Keyword or as a string.
<subsystem> – a subsystem as a string.
<area> – an area as a string.
If there are no parameters, then the list of subsystems is returned.
If there is a valid subsystem parameter, then the list of areas for that subsystem is
returned.
If the subsystem parameter is "ALL", then all the procedures of all the
areas of all the subsystems is returned. Each procedure is prefixed with
"<subsystem>:<area>:" and separated by a comma. Lists are always in priority of
desired execution.
If the area parameter is "ALL", then all the procedures of all the areas for a
specified subsystem is returned. Each procedure is prefixed with "<area>:" and
separated by a comma. Lists are always in priority of desired execution.
If the subsystem and area parameters are valid, then the list of procedures for that
subsystem and area is returned.
Returns
Examples
String of all calibration "subsystems", "areas" and/or "procedures" separated
by commas.
CALIBRATION:CATALOG? might return
"Initialization,Channel1,Channel2,System".
CALIBRATION:CATALOG? "Channel1" might return
"Dc,Adc,Clock,Align,Dac,Marker1,Marker2"
CALIBRATION:CATALOG? "ALL" might return "Initialization:Init:Calibration
Initialization,Channel1:Dc:Differential Offset,Channel1:Dc:Common
Mode,Channel1:Dc:Amplitude,Channel1:Adc:Adc
Internal,Channel1:Clock:Clock Amplitude,Channel1:Clock:Clock Offset,
Channel1:Align:Sample Point, Channel1:Dac:Speed"
CALibration:LOG? (Query Only)
This command returns a string of continuous concatenated calibration results.
The start time is recorded plus one or more <cal path>:<cal name> <result>.
This command can be issued while calibration is still in progress. Use the
CALibration:LOG:CLEar command to start a fresh log and provide additional
information.
Log results are still valid if the calibration is aborted and the calibration constants
are restored.
AWG70000 Series Programmer Manual
2-45