Download Cardio7 Operation Manual
Transcript
Cardio7 Operation Manual Printing grid may be needed when a fax paper is used instead of EKG paper for printing. For grid line will be shown for every 5mm X 5mm space and dots for every 1mm X 1mm space. If EKG paper(thermal paper) supplied by Bionet is used, Grid option should be turned off as it already contains grid on paper. Grid option could also be set on Printing Form menu. In Grid menu, use rotary key or touch screen to change grid option between ON and OFF. When Grid option is set to ON, grid will be printed out, and no grid printout for OFF. Beat Report Setup During diagnostic print, set output of the representative beat information. Beat report is added to regular printout (3ch+1rhy, 3ch+3rhy, 6ch+1rhy and 12ch rhy output), and has “Representative Beat Report’ printed on the upper part of the printout form. This Beat report has representative waveforms of each lead and shows any important parameters calculated from recordings. In Beat menu under Printing Form menu, use the rotary key or touch screen to select between ON and OFF to turn the option on or off. ON will cause Beat report to be printed and OFF to disable beat report printing. However, at 60sec and 5min recording mode, Beat Report is not printed. 38/81 Version 1.0