Download S/5 Avance Anesthesia Machine
Transcript
S/5 Avance Error Log Entry Alarm Text Condition (Basic info) Priority Source Enabling Criteria Action/Troubleshooting INSP FLOW SENSOR EEPROM FAILURE Replace insp flow sensor EEPROM cal data read failure Low AC, Vent Replace the Inspiratory Flow Sensor. LOSS OF GAS DELIVERY USER SETTINGS After regular communications has been established between the AC and the Display Computer, this alarm shall be declared if the system is in the Therapy State and the AC determines the Gas Delivery User Settings from the Display Computer arrived more than 10 seconds ago. AC Reboot system. If problem continues, check the ACB to Mixer communication LED’s (or VIB communiction LED’s). 1. If the RCV and XMT (or TXD and RXD) LED’s indicate activity, • check DU cable connections. • replace Display Controller board if problem continues. 2. If the RCV and XMT ( or TXD and RXD) LED’s indicate no activity, • check the Anesthesia Contro board connection. • replace the Anesthesia Control board if problem continues. LOSS OF VENT Vent Fail. Monitoring PARAMETER SETTINGS Only After regular communications Medium has been established between the AC and the Display Computer, this alarm shall be declared if the system is in the Therapy State and the AC determines the Ventilator Parameter Settings from the Display Computer arrived more than 10 seconds ago. AC Reboot system. If problem continues, check the ACB to Mixer communication LED’s (or VIB communiction LED’s). 1. If the RCV and XMT (or TXD and RXD) LED’s indicate activity, • check DU cable connections. • replace Display Controller board if problem continues. 2. If the RCV and XMT ( or TXD and RXD) LED’s indicate no activity, • check the Anesthesia Contro board connection. replace the Anesthesia Control board if problem continues. 7-34 11/03 1009-0357-000