Error Number You can only make a D2/Dl measurement if both
152
channels have the same number of points.
Error Number A fatal firmware error occurred before instrument
111
preset completed. Call your local Hewlett-Packard sales
and service office.
DE~KIDIJLHT I
OH NijT &TIL I D
Error Number Demodulation was selected when the analyzer was not
17
in CW tune mode. Select demodulation only after
putting the analyzer into CW time mode.
DE’$IC:E: n&. orI3 no . t. connect. r wroriq ad&-5
Error Number The device at the selected address cannot be accessed
119
by the analyzer. Verify that the device is switched on,
and check the HP-ID connection between the analyzer
and the device. Ensure that the device address
recognized by the analyzer matches the HP-ID address
set on the device itself.
[> I REI:TI:IR’$
FIJLL
Error Number There is no room left in the directory to add files.
188
Either delete files or get a new disk.
7-10 Error Messages