Download Sonicator® Plus 920
Transcript
Sonicator Plus 920 Instruction Manual — Rev. F_07/02/12 Section 4—Operating Instructions Figure 4.1—Membrane panel and touch-sensitive screen 4.1 A Note about Electrodes To ensure safe operation of the Sonicator Plus 920, follow the recommendations listed below: 1. We strongly encourage careful maintenance of the electrode system. This includes the lead wires as well as the pads themselves. Worn cables and/or poor pads (or the wrong sized pads) can have a significant impact upon treatment results. 2. Do not exceed the number of recommended uses listed on the instructions for V Trodes or other reusable self–adhesive electrodes. 3. Make sure that the entire surface of the electrode is contacting the patient. 4. Do not use moist hot packs to over the electrodes. 5. To avoid skin irritation due to high current density, do not use electrodes smaller in surface area than the 2" in diameter V Trode self-adhesive electrode (ME 2702). 6. Do not use conductive carbon electrodes with this product. 7. Do not use self-adhesive electrodes with the Direct Current mode. Use sponge type electrodes with plain tap water moistening the sponges. 8. Whenever clinically possible, utilize the largest possible pads to reduce local increases in current density. In situations where small pads are required, use the lowest stimulation intensity necessary to achieve the desired clinical results. 21