The Electrodes Window is used to test electrode impedances for all recording and patient ground and
stimulation electrodes used in the test. When a test is first started, electrodes are measured
automatically. If all electrodes impedances are satisfactory, the EMG Window will be displayed and the
EMG test started.
During the test, electrodes are continuously monitored. If an electrode lead fails or if its impedance
becomes too high to provide accurate monitoring a warning message will be displayed.
Insure electrode impedances are lower than 5K Ohms and attempt to match impedances of each channel
to less than 2K Ohms.
Each electrode is measured independently, however at least 3 electrodes, including the Patient Ground
electrode must be connected to obtain accurate readings.
Proper application of recording electrodes is essential for accurate monitoring and reduction of power
line related interference. Power line interference appears as a 60 Hz, or multiples of 60 Hz, periodic
waveform “riding” on the trace. It can be caused by either electrostatic or electromagnetic induction
from the power line to the patient or instrument. Reducing or balancing electrode impedance or moving
the patient position relative to the equipment or changing the site may reduce this artifact.
The Electrodes Window is divided into the Impedance Measurement Area and Body Graphic Area.
15.1 STIMULATION ELECTRODES MEASUREMENT AREA
The top right portion of the Impedance Measurement Area displays the stimulation electrode impedance
of those electrodes used in the test. Each box in the area displays the impedance of the positive, negative
or stimulator return electrode. Stimulation electrode impedances should be as low as possible, preferably
under 2k Ohms. Impedances above 15K Ohms are indicated by a red color. These electrodes should by
reapplied and retested.