Download Weight Controller HI-3030
Transcript
9 CHAPTER 2 Calibration the Channel you want to calibrate and repeat the C2 calibration procedures. C2 Quick Calibration From the Web Page Step 1. On the Weight Controller Home Page Click on Configuration. (See Fig. 2-8) The Configuration page appears. (See Fig. 2-9) FIG. 2-10 CALIBRATION SUB-MENU FIG. 2-8 CONFIGURATION MENU/SELECTING SETUP Step 3. Click on the Channel pull down menu you will see up to 4 channels to select. Select the scale you want calibrate by clicking on the channel number. Step 4. The Load Sensor number is a read only field. It tells you how many load sensors are connected to the instrument. Step 5. To enter the Reference Weight click in the Reference Weight field. (See Fig. 2-10) a. b. The Reference Point is the total weight that is currently on the scale. If you have nothing on the scale the Ref Point is 0. If you have 5 lbs on the scale the Ref Point is 5. Step 6. To clear the entry, move the cursor over the current reference weight which highlights the weight value. Step 7. Use your keyboard to type in the new value. In our example we entered 3.00. (See Fig. 2-10) Step 8. Click on the Do C2 Calibration button. Step 9. A page telling you that the C2 Calibration completed OK appears. (See Fig. 2-11) FIG. 2-9 CONFIGURATION PAGE Step 2. Click on Calibration. The Calibration Sub-menu appears. (See Fig. 2-10) FIG. 2-11 CAL COMPLETED OK Step 10. Click on “Back” to return to the Calibration page. Step 11. Click on “Home” to return to the Weight Controller Home page. Step 12. C2 calibration is complete.