Download CT/e Dual 3.x CONFORMANCE STATEMENT
Transcript
GE MEDICAL SYSTEMS DIR 2378622 REV 2.2 CT/E DUAL DICOM CONFORMANCE STATEMENT 6.6.2 Configurable Parameters The following fields are configurable for the DICOM Print SCU AE: Note: All configurations should be performed by a GE Field Service Engineer. • Local IP Address • Local IP Netmask • Max PDU length • The following DICOM print parameters are configurable. The valid ranges are shown in earlier sections of this conformance statement. • • • Medium Type (2000, 0030) • Film Destination (2000, 0040) • Magnification Type (2010, 0060) • Min Density (2010, 0120) • Max Density (2010, 0130) • Empty Image Density (2010, 0110) • Border Density (2010, 0100) • Configuration Information (2010, 0150) • Smoothing Type (2010, 0080) The following network timers are supported and are configurable. When these timers expire the association is terminated. The default values are: • Association Time out = 600 seconds • Session Time out = 3000 seconds (total time for open association) • N-SET Time out = 300 seconds • N-ACTION Time out = 300 seconds • N-CREATE Time out = 300 seconds • N-DELETE Time out = 300 seconds • N-GET Time out = 100 seconds Time-outs, which are set for all hosts, are configurable in dprint.cfg: *Association time-out - bi_assoc_tio Do not duplicate 95