Download CT/e Dual 3.x CONFORMANCE STATEMENT

Transcript
GE MEDICAL SYSTEMS
DIR 2378622 REV 2.2
Attribute Name
CT/E DUAL DICOM CONFORMANCE STATEMENT
Tag
Usage in MPPS Instance
Acquisition with
MWL data
Acquisition without
MWL data
Post-Processing
Modality
(0008,0060)
Value “CT” is
stored in MPPS
object
Value “CT” is
stored in MPPS
object
Value “CT” is stored
in MPPS object
Study ID
(0020,0010)
Same as exam
number
Same as exam
number
Copied from source
image
Performed Action Item
Code Sequence
(0040,0260)
Copied from SPS, Not used
if selected
Not used
Performed Series
Sequence
(0040,0340)
One item for each
series created
One item for each
series created with
post-processing
>Performing Physician's
Name
(0008,1050)
Copied from SPS, User input on the
if selected
scanner
Not sent as part of
MPPS object
>Protocol Name
(0018,1030)
The name of the
protocol selected
on the scanner
Copied from source
image
>Operator's Name
(0008,1070)
Copied from SPS, User input on the
if selected
scanner
Not sent as part of
MPPS object
> Series Instance UID
(0020,000E)
Created at the
scanner
Created at the
scanner
Created at the
scanner
> Series Description
(0008,103E)
User input on the
scanner
User input on the
scanner
Unique name
generated by
post-processing
>Retrieve AE Title
(0008,0054)
Local system AE
Title
Local system AE
Title
AE Title of the
system
>Referenced Image
Sequence
(0008,1140)
One item for each
image created
within the series
One item for each
image created
within the series
One item for each
image generated by
post-processing
>Referenced Standalone
SOP Instance Sequence
(0040,0220)
Not sent as part of Not sent as part of
MPPS object
MPPS object
Not sent as part of
MPPS object
>All other attributes from
Performed Series
Sequence
Not sent as part of Not sent as part of
MPPS object
MPPS object
Not sent as part of
MPPS object
All other attributes from
Radiation Dose Module
and Billing and Material
Code Module
Not sent as part of Not sent as part of
MPPS object
MPPS object
Not sent as part of
MPPS object
One item for each
series created
The name of the
protocol selected
on the scanner
Note 1:
•
When PPS start ( N-CREATE) message is sent, this element will have the value “INPROGRESS”
•
When PPS end ( N-SET) message is sent, this element will have either “COMPLETE” or “DISCONTINUE”
based on user selection
8.2 CT DICOM IMAGE HEADER MAPPING
Do not duplicate