Download ROTOGRAPH ROTOGRAPH-D - Villa Radiology Systems
Transcript
SERVICE MANUAL Technical features 4.3 Technical factors measuring method kVp The peak tube potential is directly measured with a non invasive kVp-meter, accuracy ±3kVp. When performing the measurement, make sure that measuring probe is completely covered by the X-ray beam. A direct measurement of the high voltage can only be carried out by specialized technicians in a suitable testing laboratory as it requires disassembling of the tube head. mA The output current is determined by measuring the voltage drop on a resistor (1kΩ, 5%) using a digital multimeter, connected to the corresponding plugs, as indicated on paragraph 7.4 (digital multimeter set to VDC 20V, 1V=1mA). t The exposure times are determined by using a timer/counter, having an accuracy of 0.1%, measuring the duration of part of the voltage applied to the primary side of the tube head, during the exposure phase. (Rev. 0) 4-7 ROTOGRAPH-D (120V)