D7000 Series - Radar Level Instrument - User's Manual
Product r
function C.1.14.0
For low dielectric constant products and when using TBF mode only. Type in
an exact value for dielectric constant, r.
If the product’s dielectric constant is not known, then leave at the default value
of 2. However, in this case expect reduced accuracy and / or skipping of the
measured value when the measuring mode switches over from TBF to direct
mode (for example, if r is set lower than the real value then the indicated level
will be higher than it actually is and vice versa).
Empty spectrum on/off
function C.1.17.0
Fixed and moving internals in the tank produce interference signals (false
reflections) that can result in incorrect measurements.
Empty spectrum configuration, provided as a quick setup mode, permits
interference signals to be identified and blanked out, by recording these signals
before tank filling and process startup.
The tank should ideally be completely empty and all moving parts switched on.
Follow the instructions as given in the device display
Wizard.
Empty spectrum filtering is not required if no interference from the tank is
expected.
On-site device calibration
Linearisation table
function C.1.22.0
For on-site calibration.
Linearisation procedure
1. Fill the tank to any given level.
2. Go to the linearisation table menu item via advanced setup > installation.
The level reading will be displayed with the possibility to enter a different
value
3. Make an appropriate reference measurement.
4. Type in the corrected value and confirm
5. Repeat procedure (50 points maximum).