Transcript
D7000 Series - Radar Level Instrument - User's Manual Product r function C.1.14.0 For low dielectric constant products and when using TBF mode only. Type in an exact value for dielectric constant, r. If the product’s dielectric constant is not known, then leave at the default value of 2. However, in this case expect reduced accuracy and / or skipping of the measured value when the measuring mode switches over from TBF to direct mode (for example, if r is set lower than the real value then the indicated level will be higher than it actually is and vice versa). Empty spectrum on/off function C.1.17.0 Fixed and moving internals in the tank produce interference signals (false reflections) that can result in incorrect measurements. Empty spectrum configuration, provided as a quick setup mode, permits interference signals to be identified and blanked out, by recording these signals before tank filling and process startup. The tank should ideally be completely empty and all moving parts switched on. Follow the instructions as given in the device display Wizard. Empty spectrum filtering is not required if no interference from the tank is expected. On-site device calibration Linearisation table function C.1.22.0 For on-site calibration. Linearisation procedure 1. Fill the tank to any given level. 2. Go to the linearisation table menu item via advanced setup > installation. The level reading will be displayed with the possibility to enter a different value 3. Make an appropriate reference measurement. 4. Type in the corrected value and confirm 5. Repeat procedure (50 points maximum). 98