Download Microcap Plus Service Manual.book
Transcript
Section 7: Disassembly Guide 7.8 REPLACING HOUSING COMPONENTS The Housing kit includes the following replaceable components: • • • • • • • • • • • Note: 7.8.1 7.8.2 7-12 Front Cover Rear Cover ON-OFF Button Upper Keypad (Alarm Silence/Alarm Silence Menu Button and Contrast/Value Change Button) Lower Keypad (Event/Home Button and Next/Menu Button) Gas Outlet SpO2 Insulation Plate External SpO2 Connector Latch Internal SpO2 Connector Lock Screws Washers When reassembling the monitor after replacing any part, be sure to follow the disassembly steps in reverse order. ON-OFF Button and Gas Outlet (Refer to Figure 7-4: Replacing CO2 Board and Housing Components) 1. Open the monitor as described in section 7.2 "Opening the Monitor Case". 2. Remove the Gas Outlet from the tubing and replace with a new Gas Outlet. 3. Remove the ON-OFF Button from the CO2 board as described in section 7.3 "Replacing the CO2 Board", and replace with a new button. Keypads (Refer to Figure 7-4: Replacing CO2 Board and Housing Components) 1. Open the monitor as described in section 7.2 "Opening the Monitor Case". 2. Remove the CO2 Board as described in section 7.3 "Replacing the CO2 Board". 3. Remove the old keypad. When replacing the new keypad, be sure the icons are facing the correct direction. (Refer to Figure 1-1: Handheld Capnograph/Pulse Oximeter.)