Download Notice
Transcript
LVL CR The LVL CR (level correct) function allows you to calibrate external path losses in the signal path. The level correct function creates a table in synthesizer memory of external path loss values versus frequency. 5 key and Invoking the LVL CR (level correct) function (pressing the 4 then the 4 5 key) causes the synthesizer to prompt you for the start frequency, stop frequency, and number of points in the level correction table. Once the start frequency, stop frequency, and number of points are entered, you are given the option of running the level correct routine. 5 (LVL CR) is pressed, the synthesizer initially When 4 5, 4 prompts you for the level correct start frequency. When 4 5, 4 5 (LVL CR) is pressed, the text COR START XXXGHZ will be displayed where XXX is the current start frequency. At this point, a new start frequency can be entered or the existing start frequency can be modied using the knob or arrow keys. The start frequency is used to determine the beginning frequency for the level correction table. The start frequency is the rst frequency point in the table. SHIFT INT LEVEL SHIFT SHIFT INT LEVEL INT LEVEL NOTE You do not have to choose a new start frequency (or stop frequency or number of points). To keep the same parameter value, press 4Hz5 (ENTER) without entering a new parameter and the next parameter in the sequence will be displayed. 6-31